Alert Codes
Understand the alert codes on the Vagal tVNS Stimulator.
A861
Bad setupImpossibly High Resistance
ErrorStimulation Cannot Start
- Clip Not Connected
- Impedance Impossibly High
- Check setup
On startup the device tries to pass a very small current to check if an ear is present. If it is not possible then A861 is displayed and the device automatically turns off
A862
Easy fixVery High Resistance
LimitedMild Stimulation Only
- Inspect Setup for source of resistance
- Clean Surfaces
- Apply More Conductive Paste
A862 occurs when there is high resistance between the electrodes.
A863
Don't worryMedium Resistance
AlmostCannot Reach Maximum
- Good Stimulation
- Not Quite Maximum
- Check Setup
A863 indicates there is some resistance preventing maximum stimulation.
A864
Cable Fault Detected
FaultShort Circuit Detected
- Contact support
- Please send photo of setup
The device has detected a problem on the connected jack. Only connect official Vagal cables and clip electrodes.
A865
RechargeEmpty Battery
EmptyPlease Recharge Battery
- Battery empty
- Connect Charging Cable
- To a 5v USB Power Source
- And Recharge Device
- Until Battery Charged
Please recharge the device using a vagal.com USB charging cable. Replacement cables can be ordered here - tvns charge cable